HIRANUMA Coulometric Karl Fischer Titrator AQ-2200A
The AQ-2200A high-end titrator is ideal for the trace level water content determination with its excellent capability and upgraded performance!
Features
Intuitive control panel display!
Current status, data and function keys are displayed on a colorful and large screen. Color touch screen is 7.5-inches wide. All helpful and necessary information is always at hand!
Browse data by PC on a LAN connection!
Result data with curves can be viewed on your PC internet browsing program without optional software.
Enhanced memory data capacity!
Store results data on the USB memory stick and the stored data can be read out on the color display.
Built-in thermal printer!
Easy paper loading. High resolution, clear printout.
Two KF channels run in parallel!
Combining one additional KF station with either a volumetric or coulometric module enables two different measurements to be performed at the same time. With both a volumetric and coulometric channel, one can measure moisture over the entire range from a few ppm to 100 % or double your sample throughput with 2 identical KF stations.
Specifications
Item | Specifications |
---|---|
Titration method | Coulometric Karl Fischer Titration |
Multitasking function | Up to two KF titration stations (Double coulometric channels or combination of coulometric and volumetric channels) |
End-point detection method | AC polarization potentiometric detection method |
Background correction | Automatic correction (selectable ON/OFF) |
Measuring range | 10 µgH2O to 99 mgH2O |
Display unit | µgH2O, ppm, % |
Accuracy | CV ≤ 0.3 % (1 g of 1 mgH2O/mL water-in-methanol measured 10 times) |
Required measurement time | FAST : 1.8 mgH2O/min (30 µgH2O/sec.) MEDIUM : 1.2 mgH2O/min (20 µgH2O/sec.) SLOW : 0.6 mgH2O/min (10 µgH2O/sec.) |
Electrolytic cell capacity | 150 mL (min.usage: approx. 100 mL) |
Memory data capacity | Internal memory: 100 data / channel USB Flash memory (≥256 MB): Max. approx. 9900 data |
GLP compliant functions | Electrolytic cell maintenance, Measurement accuracy test |
Calculation modes | 6 modes corresponding to sample status and sampling methods |
Printer | Built-in thermal printer (paper width: 58 mm) *Optional external printer: Dot impact printer PR-302B |
Touch panel screen | 7.5 inch color graphics LCD |
External input/output | RS-232C: 1 port, LAN: 1 port, USB flash memory stick: 1 port Interface for Evaporators: 1 port |
Power supply | AC100-240 V, 50/60 Hz, 100 VA |
Dimensions / Weight | Main unit: 220(W) × 400(D) × 200(H) mm / Approx. 5 kg |
* KF reagents are not included.
Options
Additional Units
TQ-2200 Coulometric KF titration station
includes; Amp. unit, Stirrer K-2000A2 and Electrolytic cell complete
TQV-2200 Volumetric KF titration station
includes; Amp. unit, Stirrer K-2000A2 and Titration cell complete