Environment Control Unit AFM5300E
The Hitachi research-grade AFM5300E offers significantly improved sensitivity, accuracy, and resolution of electromagnetic property measurements operated under high-vacuum conditions. Furthermore, it establishes a benchmark for comprehensive environmental control and is the only tool on the market affording AFM imaging in air/liquid/vacuum, a broad temperature range (-120 °C to 800 °C), magnetic field or humidity controls, as well as correlated AFM/SEM/ion milling investigations.
Features
1. In response to environment measurement needs (supports optic system in air/continuous heating and cooling)
This environment control AFM measures within environments that cannot be realized by conventional large scale AFM units. The electronic measuring instrument under high vacuum conditions enables control to the lowest limits the effects of surface water and material mapping of samples under heating and cooling conditions. The newly developed “Temperature Sweep Function” makes possible continuous material measurement until the cantilever touches the sample surface by inspecting outside the Z axis measurement area due to thermal expansion or shrinkage of the sample by environment temperature change and feedback control. (Patent no. 3857581, Patent no. 3926638)
- in air
- in a liquid
- in a vacuum
- temperature control
- humidity control support
2. Easy operation realized (integrated holder flange)
By employing the “Integrated holder flange open shut function” of unnecessary tools, together with simplifying the scanner replacement and introduction replacement of the sample, the optic axis adjustment after replacing the sample that resulted from the environment control AFM also became unnecessary. The holder does not need to be replaced when changing the measurement mode.
3 .Superior performance
Employs a "Swing Cancel" Function. The swinging part of a sample is fully reduced and the drift amount controlled. Confidence increased by polishing the basic performance is indispensible to nano-order structural analysis. Drift amount: less than 0.015 nm/sec.
4. Improves resolution of the electrical properties mode by reducing the influence of surface absorption water
When observicing electrical characteristics, resolution in some cases may drop from the effects of moisture on the sample surface or probe. Moisture, contaminants, etc. on the surface is eliminated by going to vacuum status. Observation of high resolution and high sensitivity becomes possible in the material observation mode.
Specifications
Specifications | AFM5300E | Optional Accessories |
---|---|---|
Manual Stage | XY ±2.5 mm | |
Sample Size | 25 mm (diameter) Thickness: 10 mm |
|
Scan range (Select at least one) |
20 µm × 20 µm × 1.5 µm | 100 µm × 100 µm × 15 µm 150 µm × 150 µm × 5 µm Closed Loop Scanner XY 1.5 µm |
Detection | Optical Lever (Low-coherence light) |
|
Optical Microscope *Top view (Select at least one) |
Zoom Microscope (Lens magnification: x7) Metallurgical Microscope (Lens magnification: × 5, × 20, × 50) |
|
Vibration Isolation | Floor-model | |
Temperature Control | -120°C to 300°C R.T. to 800°C In liquid: from R.T. to 60°C |
|
In Liquid | ✔ | |
Humidity Control | 20 to 80 %RH | |
Vacuum | Turbo-molecular Pump & Rotary Pump (9.9x10-5Pa) |
Citations
Topics
Article information on Hitachi technical magazine "SI NEWS"
This journal addresses a wide range variety of research papers and useful application data using Hitachi science instruments. Those authors are notable researchers and Hitachi application engineers. This is an Environment Control Unit AFM5300E
Describing basic principles and multiple function principles of Scanning Tunnel Microscope (STM), Atomic Force Microscope (AFM) etc.
This journal addresses a wide range variety of research papers and useful application data using Hitachi science instruments.
The trademark that represents our strong bond with the customer and shows our pledge to connect science and society to create new value.