Simultaneous Analysis of Water-Soluble Vitamins
Vitamins are essential nutrients, and are classified into water-soluble vitamins and fat soluble vitamins. We have simultaneously analyzed 9 water-soluble vitamin components by separation through a reverse phase column and diode array detection (DAD). By using DAD, detected peaks can be identified from absorption spectra. Thus, food and other samples containing many contaminants can be analyzed particularly effectively. Please take care to note that since vitamin C and erythorbic acid are unstable, they are easily decomposed during sample preparation or with the passage of time, and that it is difficult to obtain linearity and reproducibility. Therefore, these analysis conditions are suitable for qualitative analysis. For quantitative analysis, the use of an individual test method for each vitamin is recommended.
Simultaneous Analysis of Water-Soluble Vitamins
Standard Samples
Concentrations and Structural Formulas
Component
- Vitamin B1(thiamine)*
- Vitamin B6(pyridoxine) *
- Nicotinamide
- Vitamin B12(cyanocobalamin)
- Ascorbic glucoside
- Vitamin C (ascorbic acid)
- Erythorbic acid
- Vitamin B2(riboflavin)
- Nicotine acid
*:
hydrochloride salt used.
Concentration of 50 mg/L for each.
Standard stock/standard solution diluted using elute.
Conditions
Column | HITACHI LaChrom C18-PM (5 µm) (4.6 mmI.D. x 250 mm) |
---|---|
Elute | Phosphate buffer (pH 5.2)/CH3CN = 90/10 (v/v) (containing tetrabutylammonium) |
Flow rate | 0.8 mL/min |
Column Temperature | 40°C |
Detection | DAD 260 nm |
Injection vol. | 10 µL |
This analysis uses an ionpair reagent (tetrabutylammonium) as elute. As it is difficult to remove the ion-pair reagent completely from the column, use of a column that is exclusive for watersoluble vitamins is recommended.
System configuration
5110 Pump
5210 AutoSampler
5310 Column Oven
5420 UV-VIS Detector
Empower2 Data Processing System
Measurement examples for standard samples
Linearity
All of the calibration curves (all components in the range from 0.1 - 50 mg/L) exhibited high linearity, with r2= 0.995 or more. However, regarding Vitamin C, Erythorbic acid, and Vitamin B12, take care to note that it is rather difficult to obtain good linearity.
Analysis Example (1)
Nutritional Supplement Sample
Pretreatment of eye drops
A sample of weight 2 mg was weighed out, dissolved in elute to a volume of 10 mL, and filtered through a 0.45 µm filter.
Analysis Example (2)
Health Drink Sample
Pretreatment of the Health Drink
The sample was diluted 10-fold with elute and filtered through a 0.45 µm filter.
Results
The peaks of both the Nutritional Supplement and Health Drink obtained at 10.7 min tended to be identified as Erythorbic acid, which has a similar retention time. However, it was found the they were completely different substances via spectral confirmation. Thus, DAD, by which the spectral information was obtained, is effective for food and other samples containing many contaminants.
NOTE:
These data are an example of measurement; the individual values cannot be guaranteed.
The system is for research use only, and is not intended for any animal or human therapeutic or diagnostic use.
In order to read a PDF file, you need to have Adobe® Reader®