Analysis of Steroid Drugs

Drugs that contain a steroid skeleton include Prednisolone, Betamethasone, Dexamethasone, and Hydrocortisone, and each possesses a different duration of action and strength.
Steroid drugs providing powerful anti-inflammatory and immune suppression actions are used in the treatment of various illnesses, including allergy and inflammation ailments.
However, when administered continuously, steroid drugs can potentially trigger a variety of side-effects, such as obesity, high blood pressure, and diabetes. In view of this fact, when using steroids, treatment methods are needed that minimize side-effects through the proper adjustment of dosage, as well as methods of administration in a manner that is appropriate for a given disease.
The text below describes analyses that were performed on two types of off-the-shelf ointments.

Analysis of steroid drugs

Samples

Commercially available ointments A and B

Sample preparation

-

Standard samples

-

System configuration

5110 Pump
5210 AutoSampler
5310 Column Oven
5420 UV-VIS Detector
Empower2 Data Processing System

Conditions

Column HITACHI LaChrom C18 (5 µm) (4.6 mmI.D. x 150 mm)
Mobile phase H2O/MeOH = 40/60 (V/V)
Flow rate 1.0 mL/min
Column Temperature 40°C
Detection 254 nm
Injection vol. 10 µL

Results of standard sample assaying

-

Linearity

-

(Range of calibration: 0.1 - 20 µg/mL)

Conditions

Column HITACHI LaChrom C18 (5 µm) (4.6 mmI.D. x 150 mm)
Mobile phase H2O/MeOH = 40/60 (V/V)
Flow rate 1.0 mL/min
Column Temperature 40°C
Detection 254 nm
Injection vol. 10 µL

Results of analysis

-

As a result of analyzing two off-the-shelf ointments, hydrocortisone acetate was detected in Ointment A, and Prednisolone in Ointment B.

NOTE:
These data are an example of measurement; the individual values cannot be guaranteed.
The system is for research use only, and is not intended for any animal or human therapeutic or diagnostic use.